beautypg.com

Care and use manual, Iv. example of using omeprazole standard mix, V. ordering information – Waters MetID Small Molecule Standard Mix User Manual

Page 2: Vi. references

background image

MetID Standard Mix

2

[ CARE AND USE MANUAL ]

moderate precautions are taken and the samples, once reconstituted,
are stored properly at -10°C (and analyzed within an autosampler
set to 10°C or lower) the standard is typically stable for at least one
week. Beyond this time for performance testing and evaluation, fresh
re-preparation of the standard is recommended.

III. Reconstitution of the standard solution for testing

1. To the vial of Lyophilized Waters MetID Standard Mix, add 1 mL

of Methanol.

2. Recap and then agitate the vial contents (e.g. via shaking or

ultrasonic bath), to ensure the lyophilized powder dissolves in
the methanol.

3. This standard can be further diluted as per the UNIFI

Metabolite Identification Test Protocol, which documents
formal standard curve, PDA and UNIFI system performance
testing.

1

IV. Example of Using Omeprazole Standard Mix

Shown in figure 2 is a representative chromatogram under standard
chromatography conditions.

V. ORDERING INFORMATION

To order these products, contact your nearest subsidiary, or visit
www. waters.com and click on Order Center.

Description

Part number

MetID Standard Mix

186006550

Formic Acid

186006691

VI. References

1

Metabolite Identification Application Solution Quick Start P/N 715003759.

2

Maton, P.N., Omeprazole Review Article, N. Engl J Med 1991; 324:965-975.

Figure 2. Omeprazole Standard Mix (33.3 ng/mL) extracted
ion chromatogram acquired on an ACQUITY UPLC I-Class
coupled to a Xevo

®

G2-S QTof Mass Spectrometer System

and processed using UNIFI Scientific Information System.