[ CARE AND USE MANUAL ]
HILIC Quality Control Reference Material
II. STORAGE AND STABILITY
The standard mix comes in a Waters crimp top vial for direct injection and contains 1 mL total in 90:10 acetonitrile:100 mM ammonium
formate pH 3.0 buffer.
The standard mix is shipped at ambient temperatures. It is highly recommended that, upon receipt, the standard should be refrigerated at
4 °C for short term storage and for long term storage frozen.
III. EXAMPLE OF USING THE HILIC QCRM
Figure 2. Chromatographic Resolution using the HILIC QCRM.
Figure 3. HILIC QCRM across different particle sizes.
AU
0.00
0.05
0.0
0.5
1.0
1.5
2.0
2.5
3.0
3.5
4.0
4.5
5.0 min
HILIC QCRM
Lot# W11061301
1
2
3
4
1. Acenaphthene 2. Thymine 3. Adenine 4. Cytosine
1. Acenaphthene 2. Thymine 3. Adenine 4. Cytosine
AU
0.00
0.05
AU
0.00
0.05
AU
0.00
0.05
0.0
0.5
1.0
1.5
2.0
2.5
3.0
3.5
4.0
4.5
5.0 min
1
2
3
ACQUITY UPLC BEH Amide, 1.7 µm
2.1 x 150 mm
XBridge BEH Amide
XP, 2.5 µm
2.1 x 150 mm
XBridge BEH Amide, 3.5 µm
2.1 x 150 mm
4
Below is a reference chromatogram with conditions to help provide
an example of using the HILIC QCRM as a benchmark for system
performance.
System Information
Mobile phase
(A1) 90% acetonitrile/10% 100mM
ammonium formate pH 3.0 (v/v), (A2) not
used, (B1) not used, (B2) not used
Samples:
Acenaphthene
19.0 µg/mL
Thymine
3.7 µg/mL
Adenine
3.7 µg/mL
Cytosine
7.7 µg/mL
Detection wavelength
254 nm
Temperature
30 °C
System
ACQUITY UPLC
Weak needle wash
90/10 (v/v) MeCN/Milli-Q
®
Water
Strong needle wash
10/90 (v/v) MeCN/Milli-Q Water
Injection volume (µL)
2.0
Flow rate (mL/min)
0.5
Seal wash
20% MeCN/80% (v/v) Milli-Q water
Data acquisition
Empower
®
2 Software