VI. RECOMMENDED STORAGE

For overnight storage, continuously flush the column with the mobile phase at slow flow
(e.g., 50 µL/min for ACQUITY

®

System). If the column will not be used for several days,

store the column in 50% acetonitrile without any eluent additives such as formic acid or
TFA. Do not store the MassPREP Micro Desalting Columns in buffered eluents. Completely
seal the stored column to avoid evaporation and drying of the bed.