10 Appendix

72

10.2 Alarm messages

Overrange / underrange covers the following events:

- Probe break or short-circuit

- Measured value outside the probe measuring range

- Display overflow

Display

Cause

Fault remedy
(test/repair/replace)

- 1 9 9 9 9
(blinking!)

Underrange for the
value being displayed.

Is the medium being measured
within the range (too hot? too
cold?)

Check probe for break and probe
short-circuit.

Check the probe connection and
the terminals.

Check cable.

9 9 9 9 9
(blinking!)

Overrange for the value
being displayed.

all displays
on

Watchdog or power ON
trigger initialization
(reset).

Replace unit if initialization takes
longer than 5s.

PROF-ERR PROFIBUS error

Can be suppressed by setting the
PROFIBUS address to "0".

OPT-ERR

Hardware configuration
error

Check which option boards are