4 Electrical connection

18

- The instrument is not suitable for use in areas with

an explosion hazard (Ex areas).

Installation information on conductor cross
sections and core ferrules

Minimum
cross-
section

Maximum
cross-
section

Min.
length of
core-end
ferrule

Without core-end ferrule

0.34mm

2

2.5mm

2

10mm
(stripped)

Core-end ferrule without lip

0.25mm

2

2.5mm

2

10mm

Core end ferrule with lip
up to 1.5mm

2

0.25mm

2

1.5mm

2

10mm

Core end ferrule with lip
above 1.5mm

2

1.5mm

2

2.5mm

2

12mm

Twin ferrule with lip

0.25mm

2

1.5mm

2

12mm

.

V

Only allow qualified personnel to
carry out the electrical connection.

H

Identify the instrument version by
means of the type code.