7 Configuration

Special case:
Offset

If the deviation between measured value and
displayed value at the low and high measuring point
is identical, an offset correction is sufficient (ascent
remains unchanged). In this case, fine adjustment is
not required.

vMeasured value correction (offset), page 36

Repeated
fine adjust-
ment

Reset the fine adjustment prior to repeating it. For this
purpose, program start value (FINEADJB) and end
value (FINEADJE) are given the same value. This
automatically sets the start value to 0 and the end
value to 1.

*

Always check the start value and the end value
prior to starting fine adjustment.
Reset the fine adjustment, if they deviate from
the factory-set values 0 (FINEADJB) and 1
(FINEADJE).