INFICON XTM/2 Thin Film Deposition Monitor User Manual
Page 42
2 - 16
IP
N 07
4-
18
6S
XTM/2 Operating Manual
Table 2-1 Configuration Switch Settings
Switch 1
Test Mode
(0 = off, 1 = on)
Switch 2
Parameter Lock
(0 = off, 1 = on)
Switch 3
Beep On/Off
(0 = on, 1 = off)
Switch 4
Close Shutter on
Crystal Fail
(0 = yes, 1 = no)
Switch 5
Continue
Thickness/Timer
Accumulation Option
(0 = off, 1 = on)
With Switch 5 in the On position, pressing the OPEN button adds to the
Thickness and Time counters and does not zero previously accumulated
Thickness or Time. Pressing CLOSE freezes the Thickness and Time
displayed at the current value. The displayed Rate will not be frozen. The Zero
function works normally and can be used to zero accumulated Thickness or
Time at any time.
Switch 6
Unused
Switch 7
Unused
Switch 8
Etch Mode
(0 = off, 1 = on)
Switch 9
Displayed units MSB 00 = kÅ, 01 = µgm
Switch 10
Displayed units LSB
10 = mgm, 11 = MHz
Note: recorder function remains as Å/sec or Å in the
MHz setting.
Switch 11
Recorder function
MSB
00 = Rate, ±2000Å/sec, 1000ng/sec or 200µgm/sec
01 = Rate, ±200Å/sec, 100ng/sec or 20 µgm/sec
10 = Rate, ±20Å/sec, 10ng/sec or 2 µgm/sec
Switch 12
Recorder function
LSB
11 = Thickness, Modulus ±2000Å, 2000ng or 2000µgm
Switch 13
Recorder Output &
Display Averaging
MSB
00 = 1/4 sec
01 = 1 sec
Switch 14
Recorder Output &
Display Averaging
LSB
10 = 4 sec
11 = 16 sec
Note: Display average is always 1 second or greater.
Switch 15
Unused
Switch 16
Unused (Reserved)