2.3 Installation Guide and Schematic
Many experienced deposition monitor users will be able to fully install and use
the instrument by studying the installation schematic,
Figure 2-5
on the next
page, and the State Sequence Diagram,
.
A more systematic approach would be to start by reviewing the two figures and
then following the procedure below.
WARNING
Completely review
on safety.
All warnings in this section, as well as ones found in
other sections listed below, must be followed to ensure
the safety of the personnel operating this equipment.
1
Check for correct line voltage,
2
Verify basic unit operation by exercising it in the Test Mode,
3
Review the system interface capability as outlined in
. Be especially attentive of the special features available on the
configuration switches,
4
Wire the necessary connectors following the installation procedures in
, and
.
5
Review the front panel controls and display description per
6
Program the desired film parameter values per
and
.
7
Verify the operation of the just programmed film utilizing the Test Mode.
8
Attach the XIU (757-305-G15, G30, or G100) to an existing transducer or
install a new transducer following the guidelines of
and
.
9
Exit the Test Mode and deposit when ready.