XTM/2 Operating Manual

6.5.1 Diagnostic Procedures

The following diagnostic procedures employ the Crystal Sensor Emulator to
analyze a constant Crystal Fail message. The symptom is a Crystal Fail
message that is displayed by the Deposition Controller even after the monitor
crystal has been replaced with a new “good” monitor crystal.

NOTE: The "Unable To Auto Z" message will be displayed if the Crystal Sensor

Emulator is attached to a deposition controller and you are attempting
to use the Auto Z feature. This is to be expected and is normal.

6.5.1.1 Measurement System Diagnostic Procedure

1

Refer to

Figure 6-8 on page 6-20

. Remove the six-inch BNC cable from the

Feed-Through at point A.

2

Connect the Crystal Sensor Emulator to the 6 inch BNC cable at Point A.

Š If the XTAL Fail message disappears after approximately five seconds,

the measurement system is working properly. Re-install the six-inch
BNC cable to the Feed-Through. Go to

section 6.5.1.2

.

Š If the XTAL Fail message remains, continue at step 3.

3

Disconnect the six-inch BNC cable from the Oscillator and from the
Emulator.

4

Visually inspect the six-inch BNC cable to verify that the center pins are
seated properly.

5

Use an Ohm meter to verify the electrical connections on the six-inch BNC
cable.

Š There must be continuity (<0.2 ohms) between the center pins.

Š There must be isolation (>10 megohms) between the center pins and

the connector shield.

Š There must be continuity between the connector shields.

Replace the six-inch BNC cable if it is found to be defective and repeat Step
2 of this procedure.

6

If the six-inch BNC cable is not defective, re-connect the six-inch cable to
the oscillator and to the Crystal Sensor Emulator. If the XTAL Fail message
remains, contact INFICON's Service Department.