Despatch MIC1462 Controller Manual User Manual

Page 96

background image

92

Program Specifications


Programs:

Eight, each with free-form segments

Length of Programs:

Adjustable in the range 1 to 16 segments; programs
cascadable - maximum length 121 segments.

Segment Types:

Ramp, Dwell, Join, Repeat, or End.

Program Cycling:

Range 1 to 9999, infinite.

Delayed Start:

May be set in the range 0 to 99:59 (hours:minutes).
One setting applies to all programs.

Control:

Run, Hold, Abort, Time Base x60 (local or remote);
Select Program (local or remote); Jump to next Segment.

Start From:

Either current process variable value or controller setpoint value.

Go To:

Base Mode or Off Mode

End On:

Final Value or controller setpoint.

Deviation/Hold:

Off, below setpoint only, above setpoint only or above and below
setpoint. On ramps only, on dwells only, or on both ramps and
dwells. Auto/Hold band may be set from 0 to input span.

Time Base:

Either hours:minutes or minutes:secs (x60) preprogrammable or
may be set during Program Run.

Segment Time:

May be set in the range 0 to 99:59 (hours:minutes or
minutes:seconds).

Ramp Rate:

0 to 9999 least significant digits per hour or minute.



End of Program Output
Type:

Relay

Contact Type:

SPDT

Rating:

5A resistive @ 120/240V AC

Lifetime:

>100,000 operations @ rated voltage/current

Isolation:

Inherent


Event Outputs - Option
Type:

Relay (4)

Contact Type:

SPST

Rating:

5A resistive @ 120/240V AC

Lifetime:

>100,000 operations @ rated voltage/current.

Isolation:

Inherent


Programmability:

Each event is programmable to either OFF or ON for each segment.