U

SER

'

S

G

UIDE

_______________________________________________________________________

86 ___________________________________________________________________ M210194

EN

-A

>pulse
ON

SLOW LOW

0.0

>ptest 1000

(outputting pre-defined pressure 1000.0 hPa)

1000.000

1000.000

10000

>pulse slow low -800
ON

SLOW LOW

-800.0

>ptest 1000

1000.000

200.000

2000

>

NOTE

In case of an error in a PTB220 series barometer, there will be no
pulse output from the barometer despite an external trigger pulse.
This error handling technique ensures that a host system can not
receive erroneous pressure readings from a PTB220 series
barometer.

NOTE

If the analog output module is installed the PULSE output mode
cannot be used.