The response to the PTEST command in the serial interface consists
of the pressure reading, pressure reading corrected with defined offset
and the number of pulses to be output. See the PULSE command in
Chapter Operating modes for information on how to make the settings
of pulse output mode.

The switch SW3 may be in position ON or OFF (see Chapter
Hardware settings) when the PTEST command is being used.

Examples with output of measured pressure:

>pulse
ON

SLOW LOW

0.0

>ptest

1013.250

1013.250

10133

>pulse slow low -800
ON

SLOW LOW

-800.0

>ptest

1013.250

213.250

2133

>

Examples with output of pre-defined pressure (e.g. 1000 hPa):

>pulse
ON

SLOW LOW

0.0

>ptest 1000
1000.000

1000.000

10000

>pulse slow low -800
ON

SLOW LOW

-800.0

>ptest 1000

1000.000

200.000

2000

>

In case of an error, there will be no pulse output from the barometer
despite an external trigger pulse. This error handling technique
ensures that a host system cannot receive erroneous pressure readings
from a PTB220 series barometer.