EN

-A

EN 61326-1:1997 +Am 1:1998, Electrical equipment for
measurement, control and laboratory use - EMC requirements;
Generic environment.

Test methods:

Emission
Radiated emissions

CISPR16 class B (CISPR22 Class 22)

Immunity
Electrostatic discharge (ESD)

EN/IEC 61000-4-2

EM field

EN/IEC 61000-4-3

EFT Burst

EN/IEC 61000-4-4

Surge

EN/IEC 61000-4-5

Conducted RF

EN/IEC 61000-4-6

Options

Analog output module

Output range

0...5 V *, 4...20 mA

Resolution

4 Pa

Total accuracy (15...30 °C)

Class A

± 0.25 hPa

Class B

± 0.30 hPa

* Factory setting

Load resistance for the voltage output is 10...500 kΩ and for the
current output it is 0...500 Ω.
(The module provides a secondary barometer output and it is supplied
without a calibration certificate.)

Accessories

With the PTB220CASE, barometers can be converted into portable
units to be used as traveling and transfer standards. Please note that
the barometer must have the following features:

− 1, 2 or 3 pressure transducers

− class A accuracy

− local display

− RS232/485 serial interface

− Quick connector

19886 Gasket