R starting the continuous output, S stopping the output – Vaisala PTB220 User Manual
Page 50

U
SER
'
S
G
UIDE
_______________________________________________________________________
44 ___________________________________________________________________ M210194
EN
-A
T
ABLE
6-1
Commands used in RUN and STOP modes
Function
Command
starting output
R
stopping output
S
single message output
SEND
verification message
V
self-diagnostics
DNOS
pressure minimum/maximum tracking
TRACK
pressure limit listing
PLIM
resetting the barometer
RESET
R Starting the continuous output
R
This command is used to start continuous outputting in the STOP and
RUN modes and in interval output mode (see page 27).
Example:
>reset
PTB220 / 2.02
>r
1013.25 hPa
1013.25 hPa
1013.25 hPa
...
S Stopping the output
S
This command is used to stop the continuous outputting (activated
with power-up in RUN mode or using the RUN command).
Example:
>reset
PTB220 / 2.02
>r
1013.25 hPa
1013.25 hPa
1013.25 hPa
s