Cald storing the new date of calibration – Vaisala PTB220 User Manual

Page 70

background image

U

SER

'

S

G

UIDE

_______________________________________________________________________

64 ___________________________________________________________________ M210194

EN

-A

CALD Storing the new date of calibration

CALD yyyy-mm-dd

This command is used to store the new valid date of calibration in the
memory of the barometer.

Example:

>cald
Calibration date

????-??-??

>cald 1997-01-01
Calibration date

1997-01-01

>

Changing the status (ON/OFF) of the linear or multipoint corrections
or entering new linear or multipoint corrections will automatically
cancel the valid date of calibration.

NOTE

Always remember to return the switch SW4 to write DISABLE
position (OFF).