Barometric pressure

Operating range

Pressure range

s 50...1100 hPa, 500...1100 hPa

Temperature range

operating

-40 ... +60 ºC

with local display

0 ... +60 ºC

storage

-60 ... +60 ºC

with local display

-20 ... +60

°C

Humidity range

non-condensing

Accuracy

0.1 % R

1

class A

class B

Linearity

*

± 0.05 hPa ± 0.05 hPa ± 0.10 hPa

Hysteresis

*

± 0.02 hPa ± 0.03 hPa ± 0.03 hPa

Repeatability

*

± 0.02 hPa ± 0.03 hPa ± 0.03 hPa

Calibration uncertainty

**

± 70 ppm

± 0.07 hPa ± 0.15 hPa

Accuracy at +20 *C

**

*

±100 ppm ± 0.10 hPa ± 0.20 hPa

Temperature dependence

****

± 0.1 hPa ± 0.1 hPa

Total accuracy

± 0.15 hPa

± 0.25 hPa

Long-term stability

± 0.1 hPa / year

1

Accuracy specifications of Class A barometer in the measuring area

limited to 800 ... 1100 hPa pressure range and temperature of +20 ºC.

* Defined as the ±2 standard deviation limits of end-point non-
linearity, hysteresis error or repeatability error.

** Defined as ±2 standard deviation limits of inaccuracy of the
working standard including traceability to NIST.

± 0.20 hPa

± 0.

08 hPa

± 0.

08 hPa

± 0.20 hPa

± 0.3

0 hPa

± 0.

30 hPa

± 0.

45 hPa

0.20 hPa/year

Pressure range:

500...1100 hPa

50...1100 hPa