Barometric pressure
Operating range
Pressure range
s 50...1100 hPa, 500...1100 hPa
Temperature range
operating
-40 ... +60 ºC
with local display
0 ... +60 ºC
storage
-60 ... +60 ºC
with local display
-20 ... +60
°C
Humidity range
non-condensing
Accuracy
0.1 % R
1
class A
class B
Linearity
*
± 0.05 hPa ± 0.05 hPa ± 0.10 hPa
Hysteresis
*
± 0.02 hPa ± 0.03 hPa ± 0.03 hPa
Repeatability
*
± 0.02 hPa ± 0.03 hPa ± 0.03 hPa
Calibration uncertainty
**
± 70 ppm
± 0.07 hPa ± 0.15 hPa
Accuracy at +20 *C
**
*
±100 ppm ± 0.10 hPa ± 0.20 hPa
Temperature dependence
****
± 0.1 hPa ± 0.1 hPa
Total accuracy
± 0.15 hPa
± 0.25 hPa
Long-term stability
± 0.1 hPa / year
1
Accuracy specifications of Class A barometer in the measuring area
limited to 800 ... 1100 hPa pressure range and temperature of +20 ºC.
* Defined as the ±2 standard deviation limits of end-point non-
linearity, hysteresis error or repeatability error.
** Defined as ±2 standard deviation limits of inaccuracy of the
working standard including traceability to NIST.
± 0.20 hPa
± 0.
08 hPa
± 0.
08 hPa
± 0.20 hPa
± 0.3
0 hPa
± 0.
30 hPa
± 0.
45 hPa
0.20 hPa/year
Pressure range:
500...1100 hPa
50...1100 hPa