MPC Multipoint corrections
MPC x
where:
x = ON or OFF
The MPC command is used to activate or deactivate the multipoint
corrections.
The multipoint corrections are protected with switch SW4 (see
F
IGURE
7-1
). The switch SW4 is normally in write DISABLE position (OFF)
and the user must change it to write ENABLE position (ON) to be
able to turn the multipoint corrections on or off.
NOTE
Changing the multipoint corrections on or off will automatically
cancel the valid date of calibration of the.
The listing varies according to the number of transducers in the
barometer (see CORR command).
Example of multipoint adjustments performed on a barometer with
one transducer:
>mpc off
Multipoint adj:
OFF
>mpc on
Multipoint adjustments
ON
Reading Correction
499.720
-0.070
599.200
-0.080
698.710
-0.010
801.120
-0.010
900.610
-0.030
947.420
-0.020
1000.100
-0.040
1099.580
-0.040
>
NOTE
Always remember to return the switch SW4 to write DISABLE
position (OFF).