C

HAPTER

7__________________________________________________

ADJUSTMENT AND CALIBRATION

V

AISALA

_________________________________________________________________________ 59

LC Linear corrections

LC x

where:

x = ON or OFF

The LC command is used to activate or deactivate the linear offset or
offset/gain corrections.

The linear corrections are protected with switch SW4 (see

F

IGURE

7-1

).

The switch SW4 is normally in write DISABLE position (OFF) and
the user must change it to write ENABLE position (ON) to be able to
turn the linear corrections on or off.

NOTE

Changing the linear corrections on or off will automatically cancel the
valid date of calibration of the barometer (see CALD Storing the new
date of calibration command).

The listing varies according to the number of transducers in the
barometer (see CORR Listing linear and multipoint corrections).
Example of linear adjustments performed on a barometer with two
transducers:

>lc off
Linear adj.

:

OFF

>lc on
Linear adj.

:

ON

Reading Correction

Reading Correction

1013.250

-0.100

800.000

0.050

1013.250

-0.100

1000.000

-0.020

>

NOTE

Always remember to return the switch SW4 to write DISABLE
position (OFF).