>errs
E00 Nothing special to report
>

Complete list of error codes:

E00

Nothing special to report

E10

CPU EEPROM ackn. error

E20

CPU EEPROM csum

error

E11

P1

EEPROM ackn. error

E14

DAC EEPROM ackn. error

E21

P1

EEPROM csum error

E24

DAC EEPROM csum error

E31

P1

serial number error

E41

P1

f out of range

E51

P1

y-value out of range

E61

P1

pressure out of range

E71

P1

P difference too large

E12

P2

EEPROM ackn. error

E22

P2

EEPROM csum error

E32

P2

serial number error

E42

P2

f out of range

E52

P2

y-value out of range

E62

P2

pressure out of range

E72

P2

P difference too large

E13

P3

EEPROM ackn. error

E23

P3

EEPROM csum error

E33

P3

serial number error

E43

P3

f out of range

E53

P3

y-value out of range

E63

P3

pressure out of range

E73

P3

P difference too large

TEST n Testing transducer operation

TEST n

where n = number of the pressure transducer (1, 2 or 3)
The TEST n command is used to output various measured quantities
and calculated values that characterize the measurement performed by
a transducer.
For transducer P1, the listing looks as follows:

>test 1

204736

254733

321328

246633

231962

13.8283

11.1142

8.8108

11.4792

12.2053

0.57134

0.76660

-0.74368

0.04356

26.1

1033.553

1033.313

1033.313

1033.313

1033.313

1033.313

1033.313

-0.010

-0.007

0.010

>

For transducers P2 and P3, the listings can be obtained in a similar
way.