Test n testing transducer operation, Complete list of error codes – Vaisala PTB220 User Manual
Page 74

U
SER
'
S
G
UIDE
_______________________________________________________________________
68 ___________________________________________________________________ M210194
EN
-A
>errs
E00 Nothing special to report
>
Complete list of error codes:
E00
Nothing special to report
E10
CPU EEPROM ackn. error
E20
CPU EEPROM csum
error
E11
P1
EEPROM ackn. error
E14
DAC EEPROM ackn. error
E21
P1
EEPROM csum error
E24
DAC EEPROM csum error
E31
P1
serial number error
E41
P1
f out of range
E51
P1
y-value out of range
E61
P1
pressure out of range
E71
P1
P difference too large
E12
P2
EEPROM ackn. error
E22
P2
EEPROM csum error
E32
P2
serial number error
E42
P2
f out of range
E52
P2
y-value out of range
E62
P2
pressure out of range
E72
P2
P difference too large
E13
P3
EEPROM ackn. error
E23
P3
EEPROM csum error
E33
P3
serial number error
E43
P3
f out of range
E53
P3
y-value out of range
E63
P3
pressure out of range
E73
P3
P difference too large
TEST n Testing transducer operation
TEST n
where n = number of the pressure transducer (1, 2 or 3)
The TEST n command is used to output various measured quantities
and calculated values that characterize the measurement performed by
a transducer.
For transducer P1, the listing looks as follows:
>test 1
204736
254733
321328
246633
231962
13.8283
11.1142
8.8108
11.4792
12.2053
0.57134
0.76660
-0.74368
0.04356
26.1
1033.553
1033.313
1033.313
1033.313
1033.313
1033.313
1033.313
-0.010
-0.007
0.010
>
For transducers P2 and P3, the listings can be obtained in a similar
way.