Corr listing linear and multipoint corrections – Vaisala PTB220 User Manual

Page 63

background image

C

HAPTER

7__________________________________________________

ADJUSTMENT AND CALIBRATION

V

AISALA

_________________________________________________________________________ 57

T

ABLE

7-1

Adjustment and calibration commands (*only with the

optional analog output)

Function

Command

listing corrections

CORR

linear corrections on/off

LC ON/OFF

multipoint corrections on/off

MPC ON/OFF

entering linear corrections

LCI

entering multipoint corrections

MPCI

calibration date

CALD

entering linear corrections *

DLCI *

The linear and multipoint corrections are protected by switch SW4.
The switch SW4 is normally in write DISABLE position (OFF) and
the user must change it to write ENABLE position (ON) to be able to
enter new linear and multipoint corrections to the barometer.

ON

OFF

SW1

SW2

SW3

SW4

SW5

SW6

SW7

SW8

S1

F

IGURE

7-1

Switch SW4 in write ENABLE position (ON)

NOTE

Always remember to return the switch SW4 to write DISABLE
position (OFF).

CORR Listing linear and multipoint corrections

CORR

The CORR command is used to indicate the status of linear and
multipoint corrections. In case the status is ON, the corrections and
valid date of calibration are listed. The listing varies according to the
number of transducers in the barometer.