beautypg.com

Poll mode, Send outputting one single message – Vaisala PTB220 User Manual

Page 54

background image

U

SER

'

S

G

UIDE

_______________________________________________________________________

48 ___________________________________________________________________ M210194

EN

-A

POLL mode

T

ABLE

6-2

summarizes the commands used in the POLL mode. The

commands are not case sensitive.

T

ABLE

6-2

Commands used in RUN and STOP modes

Function

Command

single message output

SEND a

opening a barometer

OPEN a

closing a barometer

CLOSE

The POLL mode is used when several PTB220 series barometers are
connected to one RS232C serial interface for fine adjustment and
calibration in one batch. It is also used when several intelligent
transmitters are connected to a half duplex RS 485/422 serial
interface. See pages on page 38, on page 91 and on page 93.
In the POLL mode the barometer must have a specific address so that
the host systems will be able to direct the outputting command to the
particular barometer. The addressable SEND a command is used to
ask for a single message output from the barometer.
The OPEN a command can be used to open a single barometer
temporarily to STOP mode. The CLOSE command will restore the
POLL mode.

SEND Outputting one single message

SEND a

where:

a = the address of the barometer (0...99)
This command is used to output one message from the barometer in
the POLL mode (see SMODE command).
Example:
>addr 7

Address

:

7

>smode poll
Serial mode

: POLL

>close

line closed
send 7

(text invisible)

1013.25 hPa