Vaisala PTB220 User Manual
Page 25

C
HAPTER
4_____________________________________________________________
COMMISSIONING
V
AISALA
_________________________________________________________________________ 19
Examples of different ways of setting the output format:
1. To output the pressure reading.
>form
4.2 P " " UUUU #r #n
?
>send
1020.30 hPa
>
2. To output the transducer values (P1,P2,P3) separately and the
average value (P).
>form
4.2 P " " UUUU #r #n
? 4.2 P1 " " P2 " " P3 " " P " " UUU #r #n
>send
1020.31 1020.32 1020.33 1020.32 hPa
>
3. To output pressure reading followed with three-hour trend.
>form
4.2 P " " UUUU #r #n
? 4.2 P " " UUU " " 2.1 TREND " " UUU #r #n
>send
1020.30 hPa **.* hPa
(date not available)
>send
1020.30 hPa -1.2 hPa
4. To output pressure reading followed with three-hour trend and
pressure tendency.
>form
4.2 P " " UUUU #r #n
? 4.2 P " " UUU " " 2.1 TREND " " UUU " " A #r #n
>send
1020.30 hPa **.* hPa *
(date not available)
>send
1020.30 hPa -1.2 hPa 7
>send
1020.30 hPa
1.2 hPa 2
(no + sign)
5. To output the pressure from tranducer P1 followed with
temperature reading.
>form
4.2 P " " UUUU #r #n
? 4.2 P1 " " UUU " " 3.1 T1 " " UU #r #n
>send
1020.30 hPa
23.4 'C
(no + sign)
>