Reset resetting the barometer, Example – Vaisala PTB220 User Manual
Page 53
C
HAPTER
6________________________________________________________________
OPERATING
V
AISALA
_________________________________________________________________________ 47
Example:
>track off
Min/max track.:
OFF
>plim
Stab. level
:
0.50 hPa
Max
:
OFF
Min
:
OFF
Hi alarm
:
1100.00 hPa
Lo alarm
:
500.00 hPa
>track on
Min/max track.:
ON
>plim
Stab. level
:
0.50 hPa
Max
:
1020.31 hPa
Min
:
987.65 hPa
Hi alarm
:
1100.00 hPa
Lo alarm
:
500.00 hPa
>
If there are two or three pressure transducers in the barometer, the
Pdmax setting will also appear in the PLIM listing:
>plim
Stab. level
:
0.50 hPa
Max
:
OFF
Min
:
OFF
Hi alarm
:
1100.00 hPa
Lo alarm
:
500.00 hPa
Pd max
:
1.00 hPa
>
RESET Resetting the barometer
RESET
The RESET command is used to reset the barometer. All software
settings remain in the memory after reset or any power failure. The
RESET command must always be given if the serial bus settings, DIP
switch settings or the operating mode of the barometer have been
changed.
Example:
>reset
PTB220 / 2.02
>