beautypg.com

Reset resetting the barometer, Example – Vaisala PTB220 User Manual

Page 53

background image

C

HAPTER

6________________________________________________________________

OPERATING

V

AISALA

_________________________________________________________________________ 47

Example:

>track off
Min/max track.:

OFF

>plim
Stab. level

:

0.50 hPa

Max

:

OFF

Min

:

OFF

Hi alarm

:

1100.00 hPa

Lo alarm

:

500.00 hPa

>track on
Min/max track.:

ON

>plim
Stab. level

:

0.50 hPa

Max

:

1020.31 hPa

Min

:

987.65 hPa

Hi alarm

:

1100.00 hPa

Lo alarm

:

500.00 hPa

>

If there are two or three pressure transducers in the barometer, the
Pdmax setting will also appear in the PLIM listing:

>plim
Stab. level

:

0.50 hPa

Max

:

OFF

Min

:

OFF

Hi alarm

:

1100.00 hPa

Lo alarm

:

500.00 hPa

Pd max

:

1.00 hPa

>

RESET Resetting the barometer

RESET

The RESET command is used to reset the barometer. All software
settings remain in the memory after reset or any power failure. The
RESET command must always be given if the serial bus settings, DIP
switch settings or the operating mode of the barometer have been
changed.

Example:

>reset

PTB220 / 2.02
>