Diagnostic commands, Basic information on the barometer settings – Vaisala PTB220 User Manual
Page 72

U
SER
'
S
G
UIDE
_______________________________________________________________________
66 ___________________________________________________________________ M210194
EN
-A
T
ABLE
8-1
Serial interface factory settings
Baud rate
9600
Parity
even
Data bits
7
Stop bits
1
Duplex
full duplex
Diagnostic commands
The PTB220 series digital barometers respond to a set of diagnostic
commands, which are useful for analyzing the basic settings and
general operation of the barometers. The diagnostic commands are
listed in
T
ABLE
8-2
Diagnostic commands
Function
Command
basic information on settings
?
software version
VERS
serial number
SNUM CPU
outputting error messages
ERRS
testing transducer operation
TEST n
listing transducer coefficients
C n
testing pulse output mode
PTEST
? Basic information on the barometer settings
?
The ? command lists the basic information of the barometer.
>?
Software version
PTB220 / 2.02
Serial number
P1234567
Configuration
1
Linear adjustments
OFF
Multipoint adjustments
ON
Calibration date
1997-01-01
Baud Parity Data Stop Dpx 9600
E 7 1 F
Echo
ON
Sending mode
STOP
Measurement mode
NORMAL
Pulse mode
OFF
SLOW LOW
0.0
Address
0
Output interval
0 s