Vaisala PTB220 User Manual

Page 55

background image

C

HAPTER

6________________________________________________________________

OPERATING

V

AISALA

_________________________________________________________________________ 49

OPEN Setting a barometer from POLL mode to STOP mode

CLOSE Setting a barometer from STOP mode to POLL
mode

OPEN a CLOSE

where:

a = the address of the barometer (0 ... 99)

These commands are used to set a barometer temporarily to STOP
mode and back to POLL mode again. The OPEN command is useful
when several barometers are connected to one serial bus and
communication with a single barometer is needed using all commands
(except for the ? command, see Diagnostic commands on page 66).
The CLOSE command is a global command and requires no address.

The PTB220 series digital barometers must always be closed after the
POLL mode selection. The RESET command and powering-up will
close the PTB220 series barometers automatically.

Example:

>addr 7
Address

:

7

>smode poll
Serial mode

: POLL

>close

line closed

Example of opening and closing a barometer with address 7:

open 7

(text invisible)

PTB

7 line opened for operator commands

>

Any commands (except for the ? command) can now be used for the
barometer with address 7 without affecting other barometers with
different addresses connected to the same serial interface. However, if
the RESET command is given or the barometer is powered-up, the
barometer will automatically return to the POLL mode.