beautypg.com

Vaisala PTB220 User Manual

Page 27

background image

C

HAPTER

4_____________________________________________________________

COMMISSIONING

V

AISALA

_________________________________________________________________________ 21

>form
4.2 P " " UUUU #r #n
? 4.2 P1 " " P2 " " P3 " " P " " UUU " " ERR #r #n
>send
1020.30 1020.31 1020.32 1020.31 hPa 000
>send
1020.30 1022.31 1020.32 1020.31 hPa 010

(Pd max error)

>

>form
4.2 P " " UUUU #r #n
? 4.2 P1 " " P2 " " P3 " " P " " UUU " " ERR #r #n
>send
1020.30 1020.31 1020.32 1020.31 hPa 000
>send
1020.30 1022.31 1024.32 1022.31 hPa 111

(Pd max error)

>

10. To output the pressure reading and the stability state.

>form
4.2 P " " UUUU #r #n
? 4.2 P " " UUU OK #r #n
>send
1020.30 hPa OK

(good stability)

>send
1020.30 hPa

(poor stability)

>

11. To output the pressure reading and the checksums.

>form
4.2 P " " UUUU #r #n
? 4.2 P " " UUU " " CS2 #r #n
>send
1020.30 hPa BB
>

12. To output the pressure reading and the height corrected
pressure reading.

>form
4.2 P HCP #r #n
? 4.2 P " " hcp #r #n
>send
1011.12 1012.30
>