Vaisala PTB220 User Manual
Page 27

C
HAPTER
4_____________________________________________________________
COMMISSIONING
V
AISALA
_________________________________________________________________________ 21
>form
4.2 P " " UUUU #r #n
? 4.2 P1 " " P2 " " P3 " " P " " UUU " " ERR #r #n
>send
1020.30 1020.31 1020.32 1020.31 hPa 000
>send
1020.30 1022.31 1020.32 1020.31 hPa 010
(Pd max error)
>
>form
4.2 P " " UUUU #r #n
? 4.2 P1 " " P2 " " P3 " " P " " UUU " " ERR #r #n
>send
1020.30 1020.31 1020.32 1020.31 hPa 000
>send
1020.30 1022.31 1024.32 1022.31 hPa 111
(Pd max error)
>
10. To output the pressure reading and the stability state.
>form
4.2 P " " UUUU #r #n
? 4.2 P " " UUU OK #r #n
>send
1020.30 hPa OK
(good stability)
>send
1020.30 hPa
(poor stability)
>
11. To output the pressure reading and the checksums.
>form
4.2 P " " UUUU #r #n
? 4.2 P " " UUU " " CS2 #r #n
>send
1020.30 hPa BB
>
12. To output the pressure reading and the height corrected
pressure reading.
>form
4.2 P HCP #r #n
? 4.2 P " " hcp #r #n
>send
1011.12 1012.30
>