Troubleshooting – SHIMPO CAG-2000 1 Channel User Manual
Page 46

44
15. Troubleshooting
Phenomenon
Cause
Remedy
Air supply pressure is not
stable.
Set regulator source
pressure at 300 kPa or
more.
Regulator is
malfunctioning.
Overhaul or change it.
Measuring nozzle is worn. Change it.
Repeatability is not stable.
Piping and joints are
leaking.
Entry of moisture or oil
causes trouble to
instrument.
Check for leakage, and
retighten.
Clean instrument.
Air supply pressure is too
low or high.
Set regulator at 196 kPa.
Piping and joints are
leaking.
Check for leakage, and
retighten.
Gap of measuring nozzle
is too small.
Adjust to proper gap.
Zero adjustment knob is
ineffective.
Gap of measuring nozzle
is too large.
Adjust to proper gap.
Specified power is not
supplied.
Supply power of 85 to 264
VAC.
Zero position of air circuit
is improperly adjusted.
Calibrate instrument with
masters.
Instrument is in hold
measured value mode.
Bargraph display is
illuminated green or red
(dark).
Press to clear this
condition.
Bargraph display does not
operate.
Instrument is in set mode. End set mode.
Power is not supplied.
Supply power of 85 to 264
VAC.
Fuse is blown.
Change it.
Bargraph and alphanumeric
displays are not illuminated.
Power supply or internal
circuit is faulty.
Ask NIDEC TOSOK for
repair.