Technical data – KROHNE H250 M8-M9-M10 EN User Manual

Page 87

background image

TECHNICAL DATA

8

87

H250

www.krohne.com

04/2011 - 4000269302 MA H250-R02 en

H250U - Vertical installation position

Reference condition for gas measurements:

Reference condition for gas measurements:

Reference condition for gas measurements:

Reference condition for gas measurements:
The flow measurement of gases are refered to
SCFM or SCFH: Volume flow in Standard state 15°C, 1.013 bar abs. (ISO 13443)

Measuring span:

10 : 1

Declaration of flow:

Values = 100%

Water: 20°C [68°F]

Air: 20°C [68°F], 1.013 bar abs. [14.7 psia]

Flow direction

from top to bottom

EN

ASME

Cone

Water [GPH]

Air [SCFM]

Pressure loss

[psig]

DN15

½

"

K 15.1

17.2

0.99

2.57

K 15.2

29.1

1.55

2.62

K 15.3

44.9

2.48

2.65

K 15.4

68.7

3.72

2.94

K 15.5

111

6.20

3.23

K 15.6

172

9.92

4.26

K 15.7

291

17.4

7.64

K 15.8

396

24.8

12.3

DN25

1"

K 25.1

304

18.6

1.42

K 25.2

476

27.9

1.25

K 25.3

713

43.4

1.35

K 25.4

1189

74.4

1.69

K 25.5

2008

124

2.53

DN50

2"

K 55.1

2378

149

3.23

K 55.2

3963

248

3.38

K 55.3

5548

341

3.53

INFORMATION!
The oper. press. should be at least twice the pressure loss for liquids, and at least 5
times the pressure loss for gases! The specified pressure drops are valid for water and air at
maximum flow rate. Other flow ranges on request. Conversion of other media or operating data
(pressure, temperature, density, viscosity) is performed using the calculation method in
accordance with VDI /VDE Directive 3513

MA_H250_R02_en_269302_PRT.book Page 87 Tuesday, April 12, 2011 10:17 AM