ELECTRICAL CONNECTIONS

4

41

H250

www.krohne.com

04/2011 - 4000269302 MA H250-R02 en

4.6 Protection category

The measuring device meets all requirements of protection category IP

Therefore it is essential to observe the following points:

• Use only original gaskets. They must be clean and free of any damage. Defective gaskets have

to be replaced.

• The electrical cables used must be undamaged and must comply with regulations.
• The cables must be laid with a loop 3 upstream of the measuring device to prevent water

from getting into the housing.

• The cable glands 2 must be tightened.
• Close the unused cable glands using blanking plugs 1.

Indicator

Protection category

M9

IP65/67

M8

IP65

M10

IP66/67

DANGER!
After all servicing and maintenance work on the device, the specified protection category has to
be ensured again.

Figure 4-16: Cable gland

1 Use blanking plugs if no cable is routed through
2 Tighten cable gland firmly
3 Lay the cable in a loop