OPERATION

52

H250

www.krohne.com

04/2011 - 4000269302 MA H250-R02 en

1.4

TIME CONST.

Setting : 1 ... 20 seconds

Note:

The settable time constant affects the current output and the displayed

current flow rate.

It thus enables damped depiction if there is a highly variable flow rate.
If the current flow rate is polled via HART

®

communication, then the

transferred measured value is dependent on the time constant here, too.

1.5.1

RESET

COUNTER

YES - NO

1.5.2

RESET

ERROR

YES - NO

2.1

4-20mA OUT

The analogue current output can be set to fixed values in 10% steps

between 4.00...20.00 mA.

This function has no influence on binary switching outputs.

Note:

This test function is switched off in multidrop mode.

Display: "NOT AVAILABLE".

2.2

OUTPUT B1

OPEN

The function assignment in menu 3.2 is not taken into consideration here.

CLOSED

2.3

OUTPUT B2

OPEN

The function assignment in menu 3.3 is not taken into consideration here.

CLOSED

2.4

INPUT B3

Here there is a visual depiction of whether or not input B3 has a voltage of

5...30 V.

If input B3 is set to ACTIVE HI in menu 3.8, then the display shows "ON"

when the switching voltage is applied.

Note:

NO test function possible when the output is set to INACTIVE in menu 3.8.

3.1

LANGUAGE

ENGLISH

DEUTSCH

FRANCAIS

ITALIANO

ESPANOL

CESKY

POLSKI

NEDERLANDS

3.2

FUNCTION B1

INACTIVE

Output B1 is switched off.

SWITCHING

POINT

Output B1 switches at a set value depending on the current flow value.

COUNTER_LIM

Output B1 switches when the counter exceeds the counter limit value.

3.3

CONTACT B1

NC contact

Output B1 is normally closed. If an alarm situation occurs, the contact

opens.

NO CONTACT

Output B1 is normally open. If an alarm situation occurs, the contact

closes.

3.4

FUNCTION B2

INACTIVE

See FUNCTION B1

SWITCHING

POINT

See FUNCTION B1

COUNTER_LIM

See FUNCTION B1

PULSE OUTPUT

Output B2 generates pulses up to 10 Hz depending on the current flow

value.

Level Designation

Selection /
Input

Explanation