H250H - Horizontal installation position

Reference condition for gas measurements:

Reference condition for gas measurements:

Reference condition for gas measurements:

Reference condition for gas measurements:
The flow measurement of gases are refered to
Nl/h or Nm

3

/h: Volume flow in Normal state 0°C, 1.013 bar abs. (DIN 1343)

Measuring span:

10 : 1

Declaration of flow:

Values = 100%

Water: 20°C [68°F]

Air: 20°C [68°F], 1.013 bar abs. [14.7 psia]

EN

ASME

Cone

Water [l/h]

Air [Nm

3

/h]

Pressure loss

[mbar]

DN15

½

K 15.1

70

1.8

195

K 15.2

120

3

204

K 15.3

180

4.5

195

K 15.4

280

7.5

225

K 15.5

450

12

250

K 15.6

700

18

325

K 15.7

1200

30

590

K 15.8

1600

40

950

K 15.8

2400

60

1600

DN25

1"

K 25.1

1300

35

122

K 25.2

2000

50

105

K 25.3

3000

80

116

K 25.4

5000

130

145

K 25.5

8500

220

217

K 25.5

10000

260

336

DN50

2"

K 55.1

10000

260

240

K 55.2

16000

420

230

K 55.3

22000

580

220

K 55.3

34000

900

420

DN80

3"

K 85.1

25000

650

130

K 85.2

35000

950

130

K 85.2

60000

1600

290

DN100

4"

K 105.1

80000

2200

250

K 105.1

120000

3200

340

INFORMATION!
The oper. press. should be at least twice the pressure loss for liquids, and at least 5
times the pressure loss for gases! The specified pressure drops are valid for water and air at
maximum flow rate. Other flow ranges on request. Conversion of other media or operating data
(pressure, temperature, density, viscosity) is performed using the calculation method in
accordance with VDI /VDE Directive 3513