OPERATION

6

53

H250

www.krohne.com

04/2011 - 4000269302 MA H250-R02 en

3.5

CONTACT B2

NC contact

See CONTACT B1

NO CONTACT

See CONTACT B1

3.6

PULSE

DURATION

30 ms

50 ms

100ms

200 ms

500 ms

3.7

PULSE/UNIT

0.000001

Smallest scaling factor

Note:

In the basic setting, the unit of the pulse output corresponds to the flow

unit.
Example: volume flow unit is m

3

/h, so the pulse output is set to pulses /

m

3

or

mass flow unit is kg/h, so the pulse output is set to pulses / kg

999999.0

Largest scaling factor

3.8

FUNCTION B3

INACTIVE

ACTIVE HI

The internal counter is reset to zero when a positive voltage of from

5...30 VDC is applied to terminals R+ and R for at least 100 ms.

ACTIVE LO

The internal counter is reset to zero when a positive voltage of from 5...30

VDC applied to terminals R+ and R is interrupted for at least 100ms..

3.9

MULTIDROP

0…15

Multidrop mode means that the device works continuously in bus mode
via HART

®

communication (max. 15 devices in parallel).

The analogue current output is then set to a fixed value of 4.1 mA.
Measured values are transferred via HART

®

communication.

However, the display allows local read-off of the measured values.

The polling address can be set to 1...15.

Larger integer values are not permitted. If the polling address is set to 0,
then HART

®

bus mode is switched off.

The device is working in analogue mode. Current output 4...20 mA is

active.
Standard HART

®

communication remains guaranteed.

3.10

4mA CALIBR.

This menu item allows precise calibration of the current output.

The device generates a fixed current output of 4.00 mA.

If the measured value deviates from the displayed one, then the

measured value must be input.

When the menu is exited, the corrected value is saved.

3.11

20mA CALIBR.

This menu item allows precise calibration of the current output.

The device generates a fixed current output of 20.00 mA.

If the measured value deviates from the displayed one, then the

measured value must be input.

When the menu is exited, the corrected value is saved.

3.12

ALARM

CURRENT

OFF

Measured values > 100% are indicated as a current signal up to a

maximum of 22 mA.

ON

In the event of an error the current output is set to the fixed value of

22m A.

3.13

END & UNIT

The flow unit and the upper range value can be changed.

Note:

Changing from volume flow measurement to mass flow measurement is

only possible with a new calibration.

3.13.1

FLOW RATE

For a units list, see chapter 7.4 of the handbook

3.13.2

COUNTER

As standard, the unit for the counter is derived from the unit for the flow

measurement.

It can also be changed individually.

Level Designation

Selection /
Input

Explanation