6.4 Error messages
Error message
Description
Category
Remedy
NOT LINEARIZED Linearization faulty or
not activated =
measuring error
Error
Activate linearization or carry it out again
(HART® communication and linearization
software are required; the original calibration
values must be known), or send the device
back to the manufacturer for linearization.
NEW LINEARI.
TABLE BAD
Faulty or missing
data in the
linearization table =
measuring error
LINEARIZATIO
UNDER CONFIG
The device is in
linearization mode =
measuring error
Error
Complete the linearization and activate it
(HART® communication and linearization
software are required), or send the device
back to the manufacturer for linearization.
UNIT SYSTEM
CONFLICT
The unit for the
linearization flow is
incompatible with the
selected flow type
(mass/volume)
Error
Correct error, carry out linearization again if
necessary (HART® communication and
linearization software are required), or send
the device back to the manufacturer for
linearization.
TOO FEW
ENTRIES
The linearization
table has too few data
points
Error
Carry out linearization at at least 5 points
(HART™ communication and linearization
software are required), or send the device
back to the manufacturer for linearization.
NOT
MONOTONOUS
The sequence of the
linearization values is
not strictly monotonic
increasing
Error
Check linearization and/or carry it out again
(HART® communication and linearization
software are required), or send the device
back to the manufacturer for linearization.
FIRST NOT 0 %
The first flow value if
the linearization table
is not 0%
LAST NOT 100 %
The last flow value if
the linearization table
is not 100%
NO ZERO CAL OF
AO
The current output
zero point 4.00 mA is
not calibrated = poss.
measuring error in
the process control
system
Warning
Perform calibration using ammeter and menu
3.10 or using standard HART® tools/process
control system and poss. external ammeter.
Caution: during the calibration, switch the
measuring point to manual control.
NO F.SC. CAL OF
AO
The current output
100% = 20.00 mA is
not calibrated = poss.
measuring error in
the process control
system
Warning
Perform calibration using ammeter and menu
item 3.11 or using standard HART® tools and
external ammeter if necessary. Caution:
during calibration, switch the measuring point
to manual control.
NO TEMP.
COMPENSATION
The sensor
temperature
compensation of the
device is faulty or was
not carried out. =
possible measuring
error
Error
The device, together with an indication of the
error, must be sent back to the manufacturer
for checking.
OUTPUT NOT
LINEARIZED
Linearization is not
activated = measuring
error
Error
Activate linearization or carry it out again
(HART® communication and linearization
software are required; the original calibration
values must be known), or send the device
back to the manufacturer for linearization.
COUNTER LOST
Totalizer value was
reset by
error/overflow
Warning
Because the reset time is not known:
Controlled reset of the counter using menu
item 1.5.1 or using HART® tools/process
control system.