OPERATION

68

OPTISENS MAC 100

www.krohne.com

04/2011 - MA OPTISENS MAC 100 R01 en

6.7.1 Error category "Device failure" (bold "F")

INFORMATION!
Regard that in case of an error message of the category "F

FF

F" the measured values are not valid!

Messages on the display Description

Actions

F

F

F

F error in device

Error or failure (breakdown) of the complete

device; parameter or hardware error;

measurement not possible.

Group message, when one of the following

or some other severe error occurs.

F

F

F

F IO 1

Error or failure of corresponding IO;

parameter or hardware error;

measurement not possible.

Load settings (see C5.6.3: backup 1, backup

2 or factory settings). If status message still

does not disappear, replace electronic unit.

F

F

F

F IO 2

F

F

F

F parameter

Error or failure of data manager; parameter

or hardware error; measurement not

possible.

F

F

F

F configuration

Invalid configuration: display, software, bus

parameter or main software do not match

existing configuration.

This error message also appears if a

module has been added or removed, but

configuration has not been confirmed.

If device configuration has not been

changed, but message appears though:

replace electronic unit.

F

F

F

F display

Error or failure of the display; parameter or

hardware error; measurement not possible.

Replace electronic unit.

F

F

F

F sensor electronic A

(or B)

Error or failure of the electronics of

sensor A (or B); parameter or hardware

error; measurement not possible.

Check sensor and connection; if both are

okay, replace the electronic unit.

F

F

F

F sensor global A (or B)

Data error in the global data of the

electronics of sensor A (or B);

measurement with the affected sensor not

possible.

Load settings (see C5.6.3: backup 1, backup

2 or factory settings). If status message still

does not disappear, replace electronic unit.

F

F

F

F sensor local A (or B)

Data error in the local data of the

electronics of sensor A (or B);

measurement with the affected sensor not

possible.

Replace electronic unit.

F

F

F

F current output A (or B

or C)

Error or failure of the current output of

terminal A (or B or C); parameter or

hardware error; affected current output not

usable.

Check connection of current output; if it is

okay, replace the electronic unit.

F

F

F

F software user interface

Fault revealed by CRC check of operation

software.

Replace electronic unit.

F

F

F

F hardware settings

The set hardware parameters do not match

the identified hardware; a dialogue appears

in the display.

Answer queries in dialogue mode, follow

directions; replace defective electronic unit.

F

F

F

F hardware detection

Existing hardware cannot be identified;

modules are either unknown or defective.

Replace electronic unit.

F

F

F

F RAM/ROM error IO1

A RAM or ROM error is detected during the

CRC check.

F

F

F

F RAM/ROM error IO2