ELECTRICAL CONNECTIONS

32

OPTISENS MAC 100

www.krohne.com

04/2011 - MA OPTISENS MAC 100 R01 en

Required cable properties

• Maximum wire cross section: 1.5 mm

2

/ 0.06 square inch

• Minimum stripping length for wires: 8 mm / 0.31"

Note the following procedure, remove the cover and connect the cables accordingly:

• Start to remove the internal cover of the relay outputs (1) by pressing it down (2).
• Pull the cover forward (3), then upward out of the clip and remove it (2 and 4).
• Remove the cover of the earth terminal by pulling it upward 5.

i

You see the 10-pin terminal block which is fitted with a connected bridge.

Interference suppression (A = AC, B = DC)

1 Resistor, R = 47...390 Ω, see table before
2 Capacitor, C = 10...220 nF, see table before (e.g. Siemens MKC B 81 921)
3 Relay coil
4 Free-wheeling diode
5 Relay coil

Step 1: removing the cover

CAUTION!
The screw under the cover (5) is not to be used as a cable connection. Do not loosen or remove
the cover or the screw!