beautypg.com

Operation – KROHNE MAC 100 EN User Manual

Page 59

background image

OPERATION

6

59

OPTISENS MAC 100

www.krohne.com

04/2011 - MA OPTISENS MAC 100 R01 en

6.5.4 Menu D, service

This menu is password protected and contains functions to be used by service personnel only.

C5.7

C5.7

C5.7

C5.7

units

This menu groups several functions to define the units of the measuring

results. These unit are used for all values of the related unit group on the

display (measuring screen and parameters).

C5.7.1

concentration absolute

This function allows to define the absolute concentration unit used for

amperometric measurements.

Options: mg/l, µg/l, ppm (parts per million), ppb (parts per billion), free unit.

C5.7.2

Text free unit

This function allows to define a text if a free unit was chosen for the absolute

concentration (see previous function C5.7.1).

Conditions: at most three characters before and three after the slash!

C5.7.3

[g/l]*factor

This function allows to define the conversion factor if a free unit was chosen

for the absolute concentration (see previous function C5.7.1). With this factor

the measurement value in g/l is multiplied to get a value in the free unit.

C5.7.4

rel. concentration DO

This function allows to define the relative concentration unit used for the

measurement of DO (dissolved oxygen).

Options: %, ppm (parts per million), ppb (parts per billion).

C5.7.5

conc.solution

This function allows to define the concentration unit used for solution

measurement.

Options: %, ppm (parts per million), ppb (parts per billion).

C5.7.6

spec.conductivity

This function allows to define the specific conductivity unit.

Options: mS/cm, µS/cm, free unit.

C5.7.7

Text free unit

Properties comply with C5.7.2.

C5.7.8

[S/m]*factor

Properties comply with C5.7.3 except that the measurement value is [S/m].

C5.7.9

spec.resistance

This function allows to define the specific resistance unit.

Options: MOhm x cm, free unit.

C5.7.10

Text free unit

Properties comply with C5.7.2.

C5.7.11

[Ohm x m]*factor

Properties comply with C5.7.3 except that the measurement value is

[Ohm/m].

C5.7.12

temperature

This function allows to define the temperature unit.

Options: °C, °F, K.

C5.7.13

air pressure

This function allows to define the air pressure unit.

Options: Pa, kPa, bar, mbar, Torr, atm.

C5.7.14

probe current

This function allows to define the probe current unit.

Options: mA, free unit.

C5.7.15

Text free unit

Properties comply with C5.7.2.

C5.7.16

[A]*factor

Properties comply with C5.7.3 except that the measurement value is [A].

Level

Designation / function

Settings / descriptions

.book Page 59 Friday, April 1, 2011 4:59 PM