OPTISENS MAC 100
www.krohne.com
04/2011 - MA OPTISENS MAC 100 R01 en
C5.3.6
format 1.line
This function allows to define the number of digits after the decimal point
(i.e. the decimal places).
Options: automatic (adaptation carried out automatically according to the
available space) / X (= none) …X.XXXXXXXX (max. 8 digits).
C5.3.7
measurement 2.line
This function allows to define which measured value appears in the second
line of the display. The existence and structure of this menu within your
device depends on the hardware setting and the kind of sensor connected to
the process input. Refer to the sensor manual for further information!
C5.3.8
format 2.line
This function allows to define the number of digits after the decimal point
(i.e. the decimal places). The existence of this menu depends on the
hardware setting, the options comply with the options for the format of the
first line of the display (see C5.3.7).
C5.3.9
measurement 3.line
This function allows to define which measured value appears in the third line
of the display. The existence and structure of this menu within your device
depends on the hardware setting and the kind of sensor connected to the
process input. Refer to the sensor manual for further information!
C5.3.10
format 3.line
This function allows to define the number of digits after the decimal point
(i.e. the decimal places). The existence of this menu depends on the
hardware setting, the options comply with the options for the format of the
first line of the display (see C5.3.7).
C5.4
C5.4
C5.4
C5.4
2. meas. page
Refer to sub-menus of C5.3 "1. meas. page" (all submenus and options are
identical, except that they start with C5.4 or C5.4.X).
C5.5
C5.5
C5.5
C5.5
graphic page
The graphic page always shows the trend curve of the first measured value
of the first measuring page (see menu C5.3.2). With the help of the functions
in this submenu you can scale the X-axis (time) and the Y-axis.
C5.5.1
select range
This functions allows to define the scaling of the Y-axis.
Options: manual (set range in the following function C5.5.2) /
automatic (automatic depiction based on the measured values; the range
will be reset after a parameter change or a new start-up of the device only).
C5.5.2
range
If the option "manual" is selected in the previous function, you can define the
scaling of the Y-axis manually with the help of this function.
C5.5.3
time scale
This function allows to define the scaling of the X-axis (i.e. total time span of
the trend).
C5.6
C5.6
C5.6
C5.6
special functions
This menu groups several special functions.
C5.6.1
reset errors
Refer to A5 "reset".
C5.6.2
save settings
This function allows to save all settings and choose a storage place. For
further information refer to
Save settings on page 65.
C5.6.3
load settings
This function allows to load saved settings from different storage places. For
further information refer to
Load settings on page 65.
C5.6.4
password quick set
This function allows to define and activate a password for the access to the
quick setup menu.
Options: XXXX (if you enter four digits in the range of 0001...9999, then the
password protection is enabled) / 0000 (password protection disabled).
C5.6.5
password setup
This function allows to define and activate a password for the access to the
setup menu.
Options: XXXX (if you enter four digits in the range of 0001...9999, then the
password protection is enabled) / 0000 (password protection disabled).
Level
Designation / function
Settings / descriptions