Operation – KROHNE MAC 100 EN User Manual
Page 54
6
OPERATION
54
OPTISENS MAC 100
www.krohne.com
04/2011 - MA OPTISENS MAC 100 R01 en
6.5.3 Menu C, setup
C1, process input A
C2, process input B
C3, I/O (Inputs/outputs)
INFORMATION!
The converter has two process inputs, input A and input B. Each process input has an own
submenu in this main menu. Process input A is always present, i.e. there is always a board in the
interface "Pos. A" in the connection area. The interface of process input B only has a board with
the two channel converter. Be aware that the definition which kind of measurement a process
input can do is defined when ordering the device. The configuration cannot be changed later.
INFORMATION!
Note that the appearance of some sub-menus depends on the hardware setting and the used
sensor(s).
The structure of this menu within your device depends on the hardware setting and the kind of sensor connected to the
process input. Refer to the sensor manual for further information!
The structure of this menu within your device depends on the hardware setting and the kind of sensor connected to the
process input. Refer to the sensor manual for further information!
Level
Designation / function
Settings / descriptions
This menu groups all I/O functions together.
C3.1
C3.1
C3.1
C3.1
hardware
This sub-menu allows to select different settings of the output terminals.
The menus for the different terminals are identical, if they have the same
function (e.g. the menus of all status outputs are identical). Note that a
change of the configuration is completely available after restart of the device
(leave the menu accepting the changes)! The menu structure of the changed
output will be immediately available after the change. Parameters can be set
immediately as well, but simulations and test functions will be functional
after restart.
C3.1.1
terminals A
Setting for current output A, B or C. Options: current output (current output
active) / off (no function at terminals).
C3.1.2
terminals B
C3.1.3
terminals C
C3.1.4
terminals D
Setting for control input.
Options: control input (control input active) / off (no function at terminals).
C3.1.5
terminals R1
Setting for status output or limit switch (R1, R2 or R3, if existent). Options:
•
status output: status output active
•
limit switch: limit switch active
•
off: no function at terminals
C3.1.6
terminals R2
C3.1.7
terminals R3
C3.2
C3.2
C3.2
C3.2
current output A
Fine adjustment of the corresponding current output.
C3.2.1
range 0%…100%
Current range for the selected measurement (0…100%).
Range: 0.00...20 mA (condition: 0 mA ≤ 1st value ≤ 2nd value ≤ 21.5 mA)
C3.2.2
extended range
Defines the min. and max. limits (the limits defined in the previous function
are exceeded linearly up to these values).
Range: 3.5...21.5 mA (condition: 3.5 mA ≤ 1st value ≤ 2nd value ≤ 22 mA)
.book Page 54 Friday, April 1, 2011 4:59 PM