OPERATION
6
37
OPTISENS MAC 100
www.krohne.com
04/2011 - MA OPTISENS MAC 100 R01 en
6.2 Measuring mode
The measuring mode consists of four different measuring pages. You can scroll through them
and choose the desired one by pressing the keys ↑ or ↓.
6.2.1 First measuring page
After switching on the power, the display always shows the first measuring page. In the default
status there is no difference between the first measuring page of the one channel version and
the two channel version. Right beside all measuring results you see the corresponding unit at
the end of the line. Only if there is no unit available it is left out (this is only the case with the
diagnosis values).
INFORMATION!
•
Always push the keys at right angles to the front. Touching them from the side can cause
incorrect operation!
•
Each time you press a key, an "X" flashes in the upper right corner of the display.
•
After 5 minutes of inactivity, there is an automatic return to the measuring mode. In this case
the device does not save previously changed data!
INFORMATION!
•
The following sections show the measuring pages in the default status. You can change their
appearance in the menu mode.
•
After starting-up the device always shows the first measuring page. If you choose another
measuring page with the help of ↑ or ↓, the display shows this page continuously.
•
The converter is not able to detect the exhaustion of the connected electrodes!
INFORMATION!
Diagnosis values are all values which do not belong to the measured values of the sensor
inputs A and B (e.g. error messages). Also note the following items:
•
If there is enough space and the type of measuring result is not clear when looking at the
figure(s) and unit, there is also a short description (in the following named "label") at the
beginning of the lines.
•
The labels can appear on the first measuring page as well on the second measuring page.