OPERATION
6
51
OPTISENS MAC 100
www.krohne.com
04/2011 - MA OPTISENS MAC 100 R01 en
A6, analog outputs
A7...A16
Level
Designation / function
Settings / descriptions
Depending on the hardware setting this menu is applicable to the current outputs A/B/C, the limit switches R1/R2/R3
and the first line of the first measuring page.
A6.1
A6.1
A6.1
A6.1
measurement
Value used for driving the current output C. You can answer the question
"use all outputs?" with the following options: no (exits the function using the
value only of current output C) / yes (exits the function copying the value to
all analog outputs).
A6.2
A6.2
A6.2
A6.2
unit
Unit for the current output range, the class of the unit depends on the
measurement used at this current output (for further information refer to
function C5.7 "Units").
A6.3
A6.3
A6.3
A6.3
range
Range of the main current output C of the device. You can answer the
question "use at all output?" with the following options: no (exits the function
using the value only of current output C) / yes (exits the function copying the
value to all analog outputs).
A6.4
A6.4
A6.4
A6.4
time constant
Time constant of the current output C
The existence of the single menus depends on the hardware setting and the used sensor; refer to sensor manual for
further information.