6

OPERATION

56

OPTISENS MAC 100

www.krohne.com

04/2011 - MA OPTISENS MAC 100 R01 en

C3.5.3

information

Contains serial number, software version number and production date of the

circuit board.

C3.5.4

simulation

This function allows to simulate input values for the control input. For

further information refer to

Menu B, test on page 52and refer to Simulation

of measured values on page 60.

C3.6

C3.6

C3.6

C3.6

status output R1 or limit

switch R1

Fine adjustment of the corresponding status output or limit switch; the

function of the following sub-menus depend on the hardware setting.

C3.6.1

"mode" (status output) or

"measurement" (limit

switch)

Status output: this function allows to select the mode; note that the following

option list can be reduced depending on the available functions; for further

information about the messages refer to

Status messages and diagnostic

information on page 66). Options:

application error: output is activated if the errors "F

F

F

F error in device" or "F

application error" emerge

out of specification: output is activated if the errors "F

F

F

F error in device" or

"F application error" emerge, it is also activated if a warning of the
category "out of specification" is present

control input: signal from control input is forwarded to the status output

off: status output deactivated

error in device: output is activated if the error "F

F

F

F error in device" emerges

Limit switch: this function allows to define the value which activates the limit

switch. The available options depend on the hardware setting and the kind of

the sensor used, refer to the sensor manual for further information.

C3.6.2

"invert signal" (status

output) or "threshold"

(limit switch)

Status output: this function allows to define how the status output is

activated. Options: off (activated status output generates a high current) /

on (activated output generates a low current).

Limit switch: this function allows to define the switching level. The first

value defines the level, the second level defines a hysteresis.

C3.6.3

"information" (status

output) or "time constant"

(limit switch)

Status output: contains serial number, software version number and

production date of the circuit board.

Limit switch: filter for limit switch.

C3.6.4

"simulation" (status

output) or "invert signal"

(limit switch)

This function allows to simulate the status of the corresponding status input.

For further information refer to

Menu B, test on page 52and refer to

Simulation of measured values on page 60.
Limit switch: this function allows to define how the limit switch is activated.

Options: off (exceeded limit generates a high current at the output) /

on (exceeded limit generates a low current at the output).

C3.6.5

information

Only available for limit switch; contains serial number, software version

number and production date of the circuit board.

C3.6.6

simulation

This function allows to simulate the status for the corresponding limit

switch. For further information refer to

Menu B, test on page 52and refer to

Simulation of measured values on page 60.

C3.7

C3.7

C3.7

C3.7

status output R2 or limit

switch R2

Refer to submenus of status output or limit switch R1 (all submenus and

options are identical, except that they start with C3.7).

C3.8

C3.8

C3.8

C3.8

status output R2 or limit

switch R2

Refer to submenus of status output or limit switch R1 (all submenus and

options are identical, except that they start with C3.8).

Level

Designation / function

Settings / descriptions

.book Page 56 Friday, April 1, 2011 4:59 PM