B.72 [source:]voltage:protection:clear command, B.73 [source:]voltage:protection:tripped? query, B.74 [source:]function:mode? query – KEPCO ATE-DMG SERIES User Manual
Page 87: B.75 status:operation:condition query, B.76 status:opereration:enable command, B.72, Source:]vol, B.73, B.74, B.75
ATE-DMG 042314
B-17
returns maximum voltage protection value (see Table 1-2); this value is determined at the factory and
cannot be changed by the user. VOLT:PROT? MIN returns the minimum voltage protection value
(always 0). (See example, Figure B-6.)
B.72
[SOURce:]VOLTage:PROTection:CLEar COMMAND
VOLT:PROT:CLE
Syntax:
Short Form: [SOUR:]VOLT:PROT:CLE Long Form: [SOURce:]VOLTage:PROTection:CLEar
Description: Clears overvoltage condition. Output programmed to 0V; current limit programmed to small value
(approximately 1% of full scale current rating). Overvoltage Protection Tripped is reset so that
VOLT:PROT:TRIP? returns 0. Related Commands: VOLT:PROT, VOLT:PROT:TRIP?. (See example,
Figure B-6.)
B.73
[SOURce:]VOLTage:PROTection:TRIPped? QUERY
VOLT:PROT:TRIP?
Syntax:
Short Form: [SOUR:]VOLT:PROT:TRIP? Long Form: [SOURce:]VOLTage:PROTection:TRIPped?
Return Value: 0 or 1
Description: Identifies whether overvoltage condition was detected. Returns 0 if overvoltage protection tripped;
1 if overvoltage protection not tripped. Related Commands: VOLT:PROT, VOLT:PROT:CLE. (See
example, Figure B-6.)
B.74
[SOURce:]FUNCtion:MODE? QUERY
FUNC:MODE?
Syntax:
Short Form: FUNC:MODE?
Long Form: [SOURce:]FUNCtion:MODE?
Return Value: VOLT or CURR
Description: Identifies the operating mode of the power supply. VOLT = Constant Voltage mode (CV).
CURR = Constant Current mode (CC). (See example, Figure B-1.)
B.75
STATus:OPERation:CONDition QUERY
STAT:OPER:COND?
Syntax:
Short Form: STAT:OPER:COND?
Long Form: STATus:OPERation:CONDition?
Return Value:
Description: Returns the value of the Operation Condition Register (see Table B-2). The Operation Condition
Register contains unlatched real-time information about the operating conditions of the power supply.
Bit set to 1 = function enabled (active, true); bit reset to 0 = function disabled (inactive, false). (See
example, Figure B-7.)
B.76
STATus:OPEReration:ENABle COMMAND
STAT:OPER:ENAB
Syntax:
Short Form:
STAT:OPER:ENAB
Long Form:
STATus:OPERation:ENABle
Description: Sets Operation Enable Register. The Operation Enable Register is a mask for enabling specific bits
in the Operation Event Register which will cause the operation summary bit (bit 7) of the Status Byte
register to be set Bit set to 1 = function enabled (active, true); bit reset to 0 = function disabled (inac-
tive, false). The operation summary bit is the logical OR of all the enabled bits in the Operation Event
register. (See example, Figure B-7.)
TABLE B-2. OPERATION CONDITION REGISTER, OPERATION ENABLE REGISTER,
AND OPERATION EVENT REGISTER BITS
CONDITION
NU
CC
NU
CV
NU
WTG
NU
CAL
BIT
15-11
10
9
8
7 - 6
5
4 - 1
0
VALUE
32,768 -
2048
1024
512
256
128 - 64
32
16 -2
1
CAL - INTERFACE COMPUTING NEW CALIBRATION
CONSTANTS
CC -
POWER SUPPLY IN CONSTANT CURRENT MODE
CV -
POWER SUPPLY IN CONSTANT VOLTAGE MODE
NU -
NOT USED
WTG
WAIT FOR TRIGGER (SET BY INIT SUBSYSTEM)