B.72 [source:]voltage:protection:clear command, B.73 [source:]voltage:protection:tripped? query, B.74 [source:]function:mode? query – KEPCO ATE-DMG SERIES User Manual

Page 87: B.75 status:operation:condition query, B.76 status:opereration:enable command, B.72, Source:]vol, B.73, B.74, B.75

background image

ATE-DMG 042314

B-17

returns maximum voltage protection value (see Table 1-2); this value is determined at the factory and
cannot be changed by the user. VOLT:PROT? MIN returns the minimum voltage protection value
(always 0). (See example, Figure B-6.)

B.72

[SOURce:]VOLTage:PROTection:CLEar COMMAND

VOLT:PROT:CLE

Syntax:

Short Form: [SOUR:]VOLT:PROT:CLE Long Form: [SOURce:]VOLTage:PROTection:CLEar

Description: Clears overvoltage condition. Output programmed to 0V; current limit programmed to small value

(approximately 1% of full scale current rating). Overvoltage Protection Tripped is reset so that
VOLT:PROT:TRIP? returns 0. Related Commands: VOLT:PROT, VOLT:PROT:TRIP?. (See example,
Figure B-6.)

B.73

[SOURce:]VOLTage:PROTection:TRIPped? QUERY

VOLT:PROT:TRIP?

Syntax:

Short Form: [SOUR:]VOLT:PROT:TRIP? Long Form: [SOURce:]VOLTage:PROTection:TRIPped?
Return Value: 0 or 1

Description: Identifies whether overvoltage condition was detected. Returns 0 if overvoltage protection tripped;

1 if overvoltage protection not tripped. Related Commands: VOLT:PROT, VOLT:PROT:CLE. (See
example, Figure B-6.)

B.74

[SOURce:]FUNCtion:MODE? QUERY

FUNC:MODE?

Syntax:

Short Form: FUNC:MODE?

Long Form: [SOURce:]FUNCtion:MODE?

Return Value: VOLT or CURR

Description: Identifies the operating mode of the power supply. VOLT = Constant Voltage mode (CV).

CURR = Constant Current mode (CC). (See example, Figure B-1.)

B.75

STATus:OPERation:CONDition QUERY

STAT:OPER:COND?

Syntax:

Short Form: STAT:OPER:COND?

Long Form: STATus:OPERation:CONDition?

Return Value: 0 to 1313 (1 + 32 + 256 + 1024).

Description: Returns the value of the Operation Condition Register (see Table B-2). The Operation Condition

Register contains unlatched real-time information about the operating conditions of the power supply.
Bit set to 1 = function enabled (active, true); bit reset to 0 = function disabled (inactive, false). (See
example, Figure B-7.)

B.76

STATus:OPEReration:ENABle COMMAND

STAT:OPER:ENAB

Syntax:

Short Form:

STAT:OPER:ENAB 0 to 1313 (1 + 32 + 256 + 1024)

Long Form:

STATus:OPERation:ENABle 0 to 1313 (1 + 32 + 256 + 1024)

Description: Sets Operation Enable Register. The Operation Enable Register is a mask for enabling specific bits

in the Operation Event Register which will cause the operation summary bit (bit 7) of the Status Byte
register to be set Bit set to 1 = function enabled (active, true); bit reset to 0 = function disabled (inac-
tive, false). The operation summary bit is the logical OR of all the enabled bits in the Operation Event
register. (See example, Figure B-7.)

TABLE B-2. OPERATION CONDITION REGISTER, OPERATION ENABLE REGISTER,

AND OPERATION EVENT REGISTER BITS

CONDITION

NU

CC

NU

CV

NU

WTG

NU

CAL

BIT

15-11

10

9

8

7 - 6

5

4 - 1

0

VALUE

32,768 -

2048

1024

512

256

128 - 64

32

16 -2

1

CAL - INTERFACE COMPUTING NEW CALIBRATION

CONSTANTS

CC -

POWER SUPPLY IN CONSTANT CURRENT MODE

CV -

POWER SUPPLY IN CONSTANT VOLTAGE MODE

NU -

NOT USED

WTG

WAIT FOR TRIGGER (SET BY INIT SUBSYSTEM)