B.57 [source:]current:protection[:level]? query, B.58 [source:]current:protection:clear command, B.59 [source:]current:protection:tripped? query – KEPCO ATE-DMG SERIES User Manual
Page 84: B.57, B.58, B.59, B.60, B.61, B.62, B.60, b.61

B-14
ATE-DMG 042314
B.57
[SOURce:]CURRent:PROTection[:LEVel]?
QUERY
CURR:PROT?
Syntax:
Short Form: [SOUR:]CURR:PROT[:LEV]? {MIN | MAX}
Long Form: [SOURce:]CURRent:PROTection[:LEVel]? {MIN | MAX}
Return Value:
Description: Returns value representing current protection level.CURR:PROT? returns value set by
CURR:PROT. CURR:PROT?MAX returns maximum current protection value (see Table 1-2). This
value is determined at the factory and cannot be changed by the user. CURR:PROT?MIN returns the
minimum current protection value (always 0). (See example, Figure B-5.)
B.58
[SOURce:]CURRent:PROTection:CLEar COMMAND
CURR:PROT:CLE
Syntax:
Short Form: [SOUR:]CURR:PROT:CLE Long Form: [SOURce:]CURRent:PROTection:CLEar
Description: Clears overcurrent condition.Output programmed to 0V; current limit programmed to small value
(approximately 1% of full scale current rating). Overcurrent Protection Tripped is reset so that
CURR:PROT:TRIP? returns 0. Related Commands: CURR:PROT, CURR:PROT:TRIP? (See exam-
ple, Figure B-5.)
B.59
[SOURce:]CURRent:PROTection:TRIPped? QUERY
CURR:PROT:TRIP?
Syntax:
Short Form: [SOUR:]CURR:PROT:TRIP? Long Form: [SOURce:]CURRent:PROTection:TRIPped?
Return Value: 0 or 1.
Description: Identifies whether overcurrent condition was detected. 0 if overcurrent protection tripped. 1 if
overcurrent protection not tripped. Related Commands: CURR:PROT, CURR:PROT:CLE. (See exam-
ple, Figure B-5.)
B.60
[SOURce:]CURRent:[:LEVel]TRIGgered[:AMPlitude] COMMAND
CURR:TRIG
Syntax:
Short Form: [SOUR:]CURR[:LEV]:TRIG[:AMP]
Long Form: [SOURce:]CURRent[:LEVel]:TRIGgered[:AMPlitude]
Description: Programs current value to be transferred to output by *TRG commands. Actual output current
will depend on load conditions. If the value exceeds the maximum for the model being programmed,
error message -222,”Data out of range” is posted in output queue. If value exceeds CURR:LIM:HIGH
value, a value corresponding to the current limit will be programmed. Related Commands: CURR.
(See example, Figure B-1.)
B.61
[SOURce:]CURRent:[:LEVel]TRIGgered[:AMPlitude]?
QUERY
CURR:TRIG?
Syntax:
Short Form: [SOUR:]CURR[:LEV]:TRIG[:AMP]?
Long Form: [SOURce:]CURRent[:LEVel]:TRIGgered[:AMPlitude]?
Return Value:
Description: Returns the current value established by CURR:TRIG command. (See example, Figure B-1.)
B.62
[SOURce:]VOLTage[:LEVel][:IMMediate][:AMPlitude] COMMAND
VOLT
Syntax:
Short Form: [SOUR:]VOLT[:LEV][:IMM][:AMP]
Long Form: [SOURce:]VOLTage[:LEVel][:IMMediate][:AMPlitude]
Description: Sets programmed voltage level at power supply output. This command programs output voltage
to a specific value; actual output voltage will depend on load conditions. If the value exceeds the max-
imum for the model being programmed, error message -222,”Data out of range” is posted in output
queue. If value programmed exceeds the VOLT:LIM:HIGH value, a value corresponding to the voltage
limit will be programmed. Related Commands: *RCL, *SAV, VOLT:LIM:HIGH. (See example, Figure B-
1.