B.66 [source:]voltage:limit:high command, B.67 [source:]voltage:limit:high? query, B.68 [source:]voltage:[:level]mode command – KEPCO ATE-DMG SERIES User Manual
Page 85: B.63, B.64, Source:]vol, B.65, B.66, B.67, B.68

ATE-DMG 042314
B-15
B.63
[SOURce:]VOLTage[:LEVel][:IMMediate][:AMPlitude]? QUERY
VOLT?
Syntax:
Short Form: [SOUR:]VOLT[:LEV][:IMM][:AMP]? {MIN | MAX}
Long Form: [SOURce:]VOLTage[:LEVel][:IMMediate][:AMPlitude]? {MIN | MAX}
Description: Identifies programmed voltage, maximum allowable voltage, or minimum voltage (always 0).
The VOLT? query returns the programmed value of voltage. Actual output voltage will depend on load
conditions. The VOLT?MAX query returns the maximum voltage allowed for a particular model (e.g.,
25V for ATE100-10DMD). VOLT?MIN returns minimum voltage allowed for power supply (always 0).
Related Commands: VOLT. (See example, Figure B-6
B.64
[SOURce:]VOLTage:[:LEVel]TRIGgered[:AMPlitude] COMMAND
VOLT:TRIG
Syntax:
Short Form: [SOUR:]VOLT[:LEV]:TRIG[:AMP]
Long Form: [SOURce:]VOLTage[:LEVel]:TRIGgered[:AMPlitude]
Description: Programs voltage value to be transferred to output by *TRG commands. Actual output voltage
will depend on load conditions. If the value exceeds the maximum for the model being programmed,
error message -222,”Data out of range” is posted in output queue. If value exceeds VOLT:LIM:HIGH
value, a value corresponding to the voltage limit will be programmed. (See example, Figure B-1.)
B.65
[SOURce:]VOLTage:[:LEVel]TRIGgered[:AMPlitude]?
QUERY
VOLT:TRIG?
Syntax:
Short Form: [SOUR:]VOLT[:LEV]:TRIG[:AMP]?
Long Form: [SOURce:]VOLTage[:LEVel]:TRIGgered[:AMPlitude]?
Return Value:
Description: Returns value representing voltage value to be programmed by *TRG command established by
VOLT:TRIG command). (See example, Figure B-1.)
B.66
[SOURce:]VOLTage:LIMit:HIGH COMMAND
VOLT:LIM:HIGH
Syntax:
Short Form: [SOUR:]VOLT:LIM:HIGH
Long Form: [SOURce:]VOLTage:LIMit:HIGH
Description: Sets voltage limit for power supply output. If the value is out of the acceptable voltage range for the
power supply model, error message -222,”Data out of range” is posted in the output queue. If the user
tries to set an output voltage level larger than the VOLT:LIM:HIGH setting, a value corresponding to
the voltage limit will be programmed and error message -301,”Value bigger than limit” is posted in the
output queue. NOTE: If the voltage limit is changed, any previously established trigger levels higher
than the new limit must be reprogrammed to be lower than the new limit; otherwise triggers will cause
the output to exceed the new limit. Once the limit is established, the unit will not accept trigger values
higher than the programmed limit. Related Commands: VOLT. (See example, Figure B-6.)
B.67
[SOURce:]VOLTage:LIMit:HIGH?
QUERY
VOLT:LIM:HIGH?
Syntax:
Short Form: [SOUR:]VOLT:LIM:HIGH? Long Form: [SOURce:]VOLTage:LIMit:HIGH?
Return Value:
Description: Returns value representing voltage limit set by VOLT:LIM:HIGH command. Related Commands:
VOLT:LIM:HIGH, VOLT. (See example, Figure B-6.)
B.68
[SOURce:]VOLTage:[:LEVel]MODE COMMAND
VOLT:MODE
Syntax:
Short Form: [SOUR:]VOLT[:LEV]:MODE {FIX | EXT}
Long Form: [SOURce:]VOLTage[:MODE {FIXed | EXTernal}
Description: Establishes whether control of the output voltage is via keypad or GPIB commands (FIXED) or
via external analog 0 to 10V reference (EXTERNAL). Setting is stored in NVRAM and retained upon
subsequent power up.