Table 3-5. ieee 488 (gpib) bus interface functions, Ieee 488 (gpib) bus interface functions -20, Ieee 488 (gpib) bus command mode messages -20 – KEPCO ATE-DMG SERIES User Manual
Page 54: E 3-5 d
3-20
ATE-DMG 042314
TABLE 3-5. IEEE 488 (GPIB) BUS INTERFACE FUNCTIONS
FUNCTION
SUBSET
SYMBOL
COMMENTS
Source Handshake
SH1
Complete Capability (Interface can receive multiline messages)
Acceptor Handshake
AH1
Complete Capability (Interface can receive multiline messages)
Talker
T6
Basic talker, serial poll, unaddress if MLA (My Listen Address) (one-byte
address)
Listener
L4
Basic listener, unaddress if MTA (My Talk Address) (one-byte address).
Service Request
SR1
Complete Capability. The interface sets the SRQ line true if there is an
enabled service request condition.
Remote/Local
RL1
Complete capability. Interface selects either local or remote information. In
local mode the ATE-DMG executes front panel commands, but can be set
to remote mode via IEEE 488 bus. When in Remote mode all front panel
keys are disabled except LOCAL. LOCAL key can be disabled using key-
pad lockout command (see Appendix B, PAR. B.87) so that only the con-
troller or a power on condition can restore Local mode.
Parallel Poll
PP0
No Capability
Device Clear
DC1
Complete Capability. ATE-DMG accepts DCL (Device Clear) and SDC
(Selected Device Clear).
Device Trigger
DT1
Respond to *TRG and
Controller
C0
No Capability
TABLE 3-6. IEEE 488 (GPIB) BUS COMMAND MODE MESSAGES
MNEMONIC
MESSAGE
DESCRIPTION
COMMENTS
DCL
Device Clear
Received
GET
Group Execute Trigger
Received
GTL
Go To Local
Received
IFC
Interface Clear
Received
LLO
Local Lockout
Received
MLA
My Listen Address
Received
MTA
My Talk Address
Received
OTA
Other Talk Address
Received (Not Used)
RFD
Ready for Data
Received or Sent
SDC
Selected Device Clear
Received
SPD
Serial Poll Disable
Received
SPE
Serial Poll Enable
Received
SRQ
Service Request
Sent
UNL
Unlisten
Received
UNT
Untalk
Received