beautypg.com

5 running a program once, 6 stopping a program, 7 stopping a running program – KEPCO ATE-DMG SERIES User Manual

Page 51: 8 sample program, Table 3-3. sample program (model ate 25-40dmg), 16 calibration, Running a program once -17, Stopping a program -17, Stopping a running program -17, Sample program -17

background image

ATE-DMG 042314

3-17

3.4.15.5

RUNNING A PROGRAM ONCE

To set up a program to stop after running once, modify the program (see PAR. ) and go to the
last memory location to be executed and set the NEXT STEP address to 0. For example, with
the power supply in command entry status , press EDIT PROG. Enter the last memory location
you want executed (e.g. 14), then press ENTER. Press

key until LCD reads ViewVAL

Mem14 NEXT STEP

. Enter 0 and press ENTER. The program will now stop after memory

location 14 is executed. When the program stops, the power supply will continue to provide the
output specified by the last location (in the above example, location 14).

3.4.15.6

STOPPING A PROGRAM

To set up a program to stop after running once, set the NEXT STEP address of the last memory
location to be executed to 0. For example, with the power supply in command entry status ,
press EDIT PROG. Enter the last memory location you want executed (e.g. 14), then press
ENTER. Press

key until LCD reads ViewVAL Mem14 NEXT STEP . Enter 0 and press

ENTER. The program will now stop after memory location 14 is executed.

3.4.15.7

STOPPING A RUNNING PROGRAM

There are three ways to stop a program that is running:

Press CLEAR key: the program immediately stops at the step (location) that was being
executed when the key was pressed. The power supply output remains at the values
specified by that step.

Press RESET key: The program stops and power on defaults are restored (see PAR.
3.4.7).

Press OUTPUT ON/OFF key: the output is disabled (programmed to zero) and the pro-
gram immediately stops at the step (location) that was being executed when the key was
pressed. Pressing the OUTPUT ON/OFF key again restores the output to the values
specified by the step in effect when the program was stopped.

3.4.15.8

SAMPLE PROGRAM

Table 3-3 shows a sample program designed to cycle continuously; if the NEXT STEP value for
location 3 is set to 00, the program will run once and stop.

3.4.16

CALIBRATION

Refer to associated Service Manual for calibration of specific ATE-DMG models.

TABLE 3-3. SAMPLE PROGRAM (MODEL ATE 25-40DMG)

MEMORY

LOCATION

I SET

(Current)

(A)

V SET

(Voltage)

(V)

OCset

(Overcurrent

Protection)

(A)

OV set

(Overvoltage

Protection)

(V)

TIMEval

(0. to 2.5)

(Sec)

NEXT STEP

(Next location

to execute)

1

2.1

4.3

11

11

1.5

02

2

1.2

5.0

1.3

5.2

1.8

03

3

4.5

6.2

4.8

6.4

1.0

01