beautypg.com

In-Situ Win-Situ 4 / Pocket-Situ 4 Operators Manual User Manual

Page 92

background image

92

SECTION 11: FLOW-SENSE WIZARD

rev. 003 11/04

In the Field

4. Connect the TROLL 9000 to the PDA and navigate to the Flow-Sense

Wizard/Flow-Sense Data branch of the tree. Locate the template file
you created for the well you are sampling. Highlight that file in the tree
and tap Start from Template. The highlighted file will be used to
initialize the Wizard so that you don’t have to type as much on the
PDA.

5. Verify that the PDA settings for units used in low flow sampling are the

same as those of your desktop. These include “Low Flow units” and
units for sensors used.

6. Note that you will need to adjust Final Pumping Rate and other

variables. Let the Wizard run to stabilization.

7. Save this file (with real data) under a different name than the template

file you created on your PC. The Wizard’s default file name consists of
the Project name, Site name, Well I.D., and date, but you can rename
the file any way you like.

In the Office

8. Connect the computers in ActiveSync and the PocketSync utility will

automatically copy the resultant .FLO files back to the PC.

Note: In future visits to the same well you can re-use files from last time.