Priming the flow-sense wizard in the office, Which pocket-situ is installed. see, Priming the flow-sense wizard in – In-Situ Win-Situ 4 / Pocket-Situ 4 Operators Manual User Manual
Page 91: The office, At the end of this section
91
SECTION 11: FLOW-SENSE WIZARD
rev. 003 11/04
PRIMING THE FLOW-SENSE WIZARD IN THE OFFICE
To reduce the need for input on a PDA in the field, you may want to use
the following procedure to prepare the majority of Low Flow information on
your desktop PC ahead of time.
On the Desktop PC
1. Connect the TROLL 9000 you will use in the field to your desktop PC.
2. Create .FLO “template” files on desktop PC in preparation for going out
into the field, as follows:
For each well you need to sample, execute the Wizard and enter
reporting information. You may click Finish immediately (the sampled
data is unimportant), but be sure to run the Wizard for each well you
intend to visit and enter that well’s information.
When the Wizard asks if you want to save the file, give each one a
meaningful name, for example:
“Well abc template.flo”
“Well xyz template.flo”
3. On your next ActiveSync connection, In-Situ’s synchronization utility
PocketSync will prompt you for Low Flow templates you wish to
transfer to the PDA.
TIP: If Pocket
Sync does not
prompt for this,
click Send Low Flow Files
on the File menu.