SECTION 11: FLOW-SENSE WIZARD

rev. 003 11/04

6. Ensure the pump is on, then click or tap Start to begin the stabilization

readings.

STABILIZATION READINGS

The next screen shows for each parameter:

Reading: the current reading
Change - Val: difference between the maxium and minimum

over the last 3 readings OR

Change - %: difference between the current reading and the

third reading back (two readings ago) expressed as a
percentage

Target specified earlier in the Wizard (shown for Values)

To view this data in graphical form, click Graph.

Time the Wizard started

Elapsed time

No. of readings taken

Countdown to next
reading

TIP: In any view,
the Back button
may be used to

change the input on any
earlier screen. The program
will pause, then resume
when you are ready.