STARTING A TEST
A test set up for a Scheduled start will start by itself, assuming the device
clock is correctly set. To start a Manual start test, proceed as follows:

1. In the Navigation tree, select a pending manual test

.

2. Click or tap Start. You will be asked to confirm your selection.

3. After you confirm, the test starts and the display is updated.

If a test is running in the device, the Start button is not available. You
must stop the running test before you can start a new one. See
“Stopping a Test” below.

STOPPING A TEST

The end of a test cannot be programmed in advance. In most cases, the
Stop button is the only way to stop a running test. Exceptions: A test stops
automatically when any of the following occurs:

• when another test starts
• if the device memory fills up
• if power is lost.

1. In the Navigation tree, select a running test

.

2. Click or tap Stop. You will be asked to confirm your selection.

3. After you confirm, the test stops and the display is updated.

TIP: Once a test
starts, the following
actions will not be

available: Add, Delete, Edit
(device, parameter, test),
Calibrate.